Advanced Thermo TVS-500EX

Discontinued Model

Specifications

TVS-500EX Specifications

Measurement range

-40 to 500°C:Standard
 to 2000°C (with optional high temperature filter)

Temperature resolution

Better than 0.05°C with Averaging

Accuracy

±2°C or ±2% *1

Frame time

1/60 seconds

Detector

Uncooled FPA, 320(H) x 240(V) VOx microbolometer

Wavelength

8 to 14μm

FOV

19.4°(H)x14.6°(V) (with standard 22mm lens)

Spatial resolution

1.07mrad

Measurement distance

30cm  to ∞

Effective pixels

320(H)x240(V)

Display

3.5" semi-transmissive color LCD monitor,
Inverted display

Visible camera

640x480

Functions

Multi point temperature display

5 points

Multi point emissivity correction

5 points (Back Calculation)

Mixing Image Display

Mixing of thermal image and visible image display with sequential ration change.

Max/min temperature position tracking

Yes

Auto temperature tracking

AS (auto sense), AP (auto point), PPM mode, R&D mode, Medical mode

Digital zoom

x2, x4 (scroll display, area designation possible)

Freeze

On/Off

Isotherm function

Yes

Grid display

Yes

Alarm function

Displayed by setting the temperature
(single/continuous)

 

Signal output

Yes

Image recording

Interval recording with alarm as a trigger

Color palette

Rainbow/hotiron/ironbar/whiteblack/black white/contrast

Image improvement function

Averaging process

Other functions

Emissivity correction, day and time display, battery volume, memo, last memory

CF Card

Data storage

300/64MB. Up to 512MB.
Realtime memory (option) 400 frames (standard), 1600 frames (reduced size) *2

Mode

CF format, delete, create directory, file name input, one shot recording, interval recording, one shot playback, gallery display, file name display

File format

IRI, BMP (thermal image)/JPG (visible image)/BMP (mixing image)/LOG
Simultaneous recording of thermal image, visible image, and mixing image.

Interval recording

3S to 23H59m59s (1/60s to  when using optional realtime memory)

Temperature unit

°C/°F/K

Video output

NTSC or PAL

PC Interface

RS-232C,USB, IEEE1394(option) *2

Power

Battery (Li-ion), AC adapter, Long life battery (option)

Power consumption

14W

Battery run time

Approx 2H, approx 4H (when using optional long life battery)

Operational temperature/humidity

-10 to 50°C/RH80% or less (no condensation)

Dimensions

140(H)x140(W)x226(D)mm including protrusions

Weight

1.9Kg (without battery)

Environmental protection

IP54

Shock

30G JISC0041/IEC60068-2-27

Vibration

3G JISC0040/IEC60068-2-6

*1Accuracy applies from -20 to 300°C

*2USB, IEEE and realtime memory are exclusive to each other

Precautions for use of this product

  • Please make sure to read the instruction manual carefully before use so that the equipment can be used Safely.
  • Do not leave equipment in a place where there is water, moisture, steam, smoke, etc.
    It may cause fire, electric shock or failure of the equipment
  • This company shall not be liable for any incidental damages (loss of business profit, change or loss of data, etc.) caused by the use or non-availability of this product.
  • This company shall not be liable for any damages caused by malfunctions due to connection with other equipment or with equipment containing software developed by others.
  • This company shall not be liable for any damages caused by using the product in a way other than those explained in the operation manual.
  • The specifications and functions described in this brochure may be changed without notice for improvement.
  • Company names and product names appearing in this brochures are trade names and trademarks of those companies.
  • Windows is a trademark of Microsoft Corporation of the USA registered in the USA and other countries.
  • Since the product contains an item under export control, delivery is subject to necessary export licenses by the authorities. It is strictly regulated to export the product to certain area.
  • In case of retransfer, resale and/or reexport of the product, prior authorization by the authorities is required.

Measurement distance and FOV(with standard 22mm lens)

Measurement distance and FOV

Operational environment for the USB link software "Advanced Package"

OS compatibility

OS: Microsoft Windows 2000 SP-4, Microsoft Windows XP SP-2

PC specifications

  • PC-AT compatible
  • The above OS shall be preinstalled and USB2.0 interface shall be equipped as standard.
  • Pentium M/1.7GHz or better (Pentium 4/2.6GHz or better is recommended)
  • RAM: 512MB or more (1GB or more is recommended)
  • HDD: A blank space of approximately 1GB or more is required in the HDD. When recording thermal image at 60Fps and visible image simultaneously, 7200rpm or faster 3.5FDD will be required.
  • Displayed color: 16 bit color (65536 colors) or more
  • Resolution: XGA (1024x768) or better

This software is dedicated for TVS-500EX, and it cannot be used with other TVS series products. And this software is always sold with the hardware and not sold alone.
If the USB interface of TVS-500EX is changed to IEEE1394 interface, this software cannot be used.

Standard configuration

  • TVS-500EX
    (with 22mm standard lens)1ea
  • AC adapter
  • Battery pack
  • Battery charger
  • CF card
  • Card adapter
  • Neck strap
  • RS-232C cable
  • Operation manual
  • Pointer & Lighat
  • USB cable
  • CD for application software installation

To Top of  Page

Infrared Thermal Imaging Camera
Products
Micro Joining Equipment
Infrared Thermal Imaging Camera
Support
Infrared Thermal Imaging Camera
Intelligent Projector
Company Profile
Company's Outline
History
Offices
Contact Us
Micro Joining Equipment
Infrared Thermal Imaging Camera
Multi-Layered Board (product-mlb@ml.avio.co.jp)
Radiation Hardened POL DC/DC CONVERTER (product-me@ml.avio.co.jp)