Measuring range |
|
TH9100WB |
TH9100WBG |
Range 1: |
-40 to 120°C |
Range 2: |
0 to 500°C |
Range 3: |
200 to 2000°C
(Off-spec at 200 to 400°C)
(Temp uniformity not specified at 1500 to 2000°C) |
200 to 2000°C
(Off-spec at 200 to 400°C but can be displayed)
(Temp uniformity not specified at 1500 to 2000°C) |
Resolution |
Range 1: |
0.1°C (at 30°C 60Hz) |
0.08°C (at 30°C 60Hz) |
Accuracy |
Range 1 & 2: |
±2°C or ±2% or reading whichever is greater |
Range 3: |
Within ±32°C
(Average of 10x10 pixels at center) |
Within ±36°C
(Average of 10x10 pixels at center) |
Detector |
Uncooled focal plane array (microbolometer) |
Spectral range |
Range 1 & 2: |
8 to 14µm |
Range 3: |
3.8µm |
7.7µm |
I.F.O.V. |
1.2mrad |
Focusing range |
30cm to infinity |
Field of view |
21.7°(H) x 16.4°(V) |
Frame time |
60 frames/sec |
Display |
View finder and 3.5-inch LCD monitor with auto switch |
Thermal image pixels |
320 (H) x 240 (V) pixels |
A/D resolution |
14 bits |
Measuring functions |
Run/Freeze |
S/N improvement |
Σ2, Σ8, Σ16, Σ32, Σ64 and spatial filter ON/OFF |
Alarm |
Screen display and alarm sound (ON/OFF) |
Interval measurement |
Recording on built-in real-time memory : 1/60 to 3600 sec interval
Recording on memory card : 5 to 3600 sec interval (thermal image)
30 to 3600 sec. (thermal visual images)
Trigger function provided |
Emissivity correction |
0.10 to 1.00 (at 0.01 step), Emissivity table provided |
Env. temp. correction |
Provided (including interval NUC) |
User setup |
Pre-registration of environmental setup (max. 10 setups) |
Background comp. |
Provided |
Auto functions |
Full automatic (level, sense, focus)
Level trace, auto-gain control |
Display functions |
Thermal/visual fusion image display
Display color : color/monochrome, positive/negative
Gradation : 16, 32, 64, 128, 256
Color pallette : rainbow, brightness, shine, hot-iron,
medical, fine
Isothermal band display : max. 4 bands
Thumbnail display : 12 thermal images replay
Multi-sense display, Battery life indicator
Line-profile : X, Y line profile (waveform display)
Multilingual menu |
Image processing functions |
Variable level/sense
Multi-point temperature display (10pts)
Multi-point emissivity display (10pts)
Δt display
Max/Min (peak hold) temperature display
Alarm (full screen or specified box)
Digital zoom : 2, 4 times (Run/Freeze)
Box setting (max. 5 boxes) |
Annotation
|
Text and voice annotation (30 sec per image) |
Storage device |
Compact flash memory card for;
Thermal image in SIT or BMP file format
Visual image in SIT or JPEG file format
Thermal/visual fusion image in BMP file format |
Movie recording
|
Real-time memory : 1664 images (max. 60Hz) |
Video signal output |
NTSC/PAL, composite video signal, S-video |
Interface |
IEEE1394, RS-232C |
Operating temp/humidity |
−15 to 50°C, 90% RH or less (not condensed) |
Storage temp/humidity |
−40 to 70°C, 90% RH or less (not condensed) |
Power supply |
AC adaptor : 100V to 240V, DC 7.2V (nominal) |
Power consumption |
Approx. 6W (typical) |
Shock and vibration |
294m/sec2(IEC60068-2-27), 29.4m/sec2(IEC60068-2-6) |
Environmental protection |
IP54 (IEC60529) |
Dimensions |
Approx.108 (W) x 113 (H) x 189 (D) mm
(excluding projection and protective shield) |
Weight |
Approx. 1.7kg (excluding protective shield and protective window) |
Standard accessories |
AC adaptor, battery pack (2pcs), battery charger,
compact flash memory card, grip belt, neck strap,